St. Jude Medical, Inc. : St. Jude Medical Comments on MAUDE Database Report about Durata Lead
06/22/2012| 03:50pm US/Eastern

Recommend:
Analysis of lead from Florida hospital shows no inside-out abrasion in
Durata product
St. Jude Medical, Inc. (NYSE:STJ), a global medical device company,
today issued the following statement:
Through our investigation of the MAUDE database report submitted to the
FDA on May 2, 2012, and information provided to the company by the FDA,
including model number, implant and event dates, St. Jude Medical has
identified a single Durataź lead that matches the available information.
The FDA has confirmed that the serial number of this lead matches the
serial number from the MAUDE report. Our analysis of the lead indicates
it was damaged due to external abrasion, which is not the same as the
inside-out abrasion that has been seen in our previous generation Riataź
leads. Simultaneous to this announcement, the company has submitted a
medical device report (MDR) to the FDA with our additional findings.
The identified lead was explanted at a Florida hospital, and the
hospital has been very helpful in allowing the company to review the
details of this case. The hospital has provided us with static and
dynamic fluoroscopy and photographs of the extracted lead. The hospital
also allowed us to inspect the lead at the hospital. One of our senior
lead engineers traveled to the hospital to microscopically analyze the
lead under the hospital's supervision.
Through our investigation, we have identified that the patient had an
additional defibrillation lead that had been capped (and was no longer
being used). In the fluoroscopic images, the two leads cross in the
region mentioned in the filed MAUDE report. Based upon physical
examination, our analysis indicates the damage to the Durata lead is
consistent with external abrasion from contact with a calcified, or
hardened, heart valve or possibly from lead-to-lead contact. External
abrasion is a known cause of failure across all cardiac leads in the
industry, which is different from the inside-out abrasion seen with
externalized conductors observed in some Riata leads.
We recognize the importance of providing physicians with up-to-date and
accurate information in a timely and responsible manner so that they can
make informed patient care decisions. This case highlights the
importance of appropriate analysis in determining the root cause of lead
damage. It is our desire to work closely with physicians to understand
the circumstances surrounding a device malfunction. We encourage
physicians to report any case of lead failure to the manufacturer for
further inspection and analysis to best ensure we are able to validate
and communicate information in the interest of patient safety.
About St. Jude Medical
St. Jude Medical develops medical technology and services that focus on
putting more control into the hands of those who treat cardiac,
neurological and chronic pain patients worldwide. The company is
dedicated to advancing the practice of medicine by reducing risk
wherever possible and contributing to successful outcomes for every
patient. St. Jude Medical is headquartered in St. Paul, Minn. and has
four major focus areas that include: cardiac rhythm management, atrial
fibrillation, cardiovascular and neuromodulation. For more information,
please visit sjm.com.
Forward-Looking Statements
This news release contains forward-looking statements within the meaning
of the Private Securities Litigation Reform Act of 1995 that involve
risks and uncertainties. Such forward-looking statements include the
expectations, plans and prospects for the Company, including potential
clinical successes, anticipated regulatory approvals and future product
launches, and projected revenues, margins, earnings and market shares.
The statements made by the Company are based upon management's current
expectations and are subject to certain risks and uncertainties that
could cause actual results to differ materially from those described in
the forward-looking statements. These risks and uncertainties include
market conditions and other factors beyond the Company's control and the
risk factors and other cautionary statements described in the Company's
filings with the SEC, including those described in the Risk Factors and
Cautionary Statements sections of the Company's Annual Report on Form
10-K for the fiscal year ended December 31, 2011 and Quarterly Report on
Form 10-Q for the fiscal quarter ended March 31, 2012. The Company does
not intend to update these statements and undertakes no duty to any
person to provide any such update under any circumstance.

St. Jude Medical, Inc.
J.C. Weigelt
Investor Relations
651-756-4347
jweigelt@sjm.com
or
Amy
Jo Meyer
Media Relations
651-756-3029
ameyer@sjm.com
© Business Wire 2012
Recommend :