1. Overview

JMS-Q1600GC UltraQuad™ SQ-Zeta

Helium (He), which is widely used as a carrier gas for GC, may face problems such as temporary price increases or unstable supply conditions due to various reasons, and when supply delays occur, it is necessary to consider using another type of carrier gas as an alternative. Hydrogen and nitrogen are the main alternative gases being considered. Hydrogen, in particular, has a wide linear velocity range for optimum separation and is suitable as a carrier gas for GC. In this report, di(2-ethylhexyl)phthalate, dichloroacetonitrile, and chloral hydrate, which are water quality control target items in water quality testing, were measured using hydrogen carrier gas by the same column. (For He, see MSTips No. 325.) As a result, good linearity of the calibration curve and reproducibility at the lower limit of quantification were obtained for all compounds, and are presented in this report.

1.1. Measurement Condition

The measurements were performed using a gas chromatograph quadrupole mass spectrometer "JMS-Q1600GC UltraQuad™ SQ-Zeta". The measurement conditions are shown in Table 1. As mentioned above, the column used was the same, and each compound was measured by changing the GC conditions. The monitor ions for SIM described in the Japanese "Notification Law" were used.

Table. 1 Measurement Condition of Each Compound

Parameter Value
Di(2-ethylhexyl)phthalate Dichloroacetonitrile & Chloral hydrate
GC Oven temp. 50°C (2°Cmin)→20°C/min→180°C(0 min)→
5°C/min→260°C (10.5 min)→10°C/min→
280°C (5 min), Total 42 min
35°C (3.5 min)→15°C/min→100°C (0 min)→
20°C/min→250°C (3 min),
Total 18.3 min
Column flow (Hydrogen) 1.3 mL/min 1.5 mL/min
Injection mode Splitless, Purge time 1 min PulsedSplitless, Purge time 0.4 min
Injection volume 1 uL 2 uL
Column GL Sciences Inc. InertCap 1 MS, 30 m x 0.25 mm id, 1 μm film thickness
Inlet temp. 250°C
MS Interface temp. 250°C
Ion source temp. 250°C
Ionization EI (70 eV, 50 μA)
Acquisition mode SIM
2. Di(2-ethylhexyl)phthalate 2.1. Measurement

The test solution was diluted with n-hexane to achieve concentrations of 5, 10, 15, and 20 μg/L of di(2-ethylhexyl)phthalate in the test water before pretreatment. Phenanthrene-D10 was added as an internal standard at a concentration of 250 μg/L.

2.2 Result

The calibration curve for di(2-ethylhexyl)phthalate is shown in Figure 1. The coefficient of correlation for the relative calibration curve was more than 0.999. SIM chromatogram and the coefficient of variation of the quantitation value for the 5 μg/L sample measured continuously at n=5 are shown in Figure 2 and Table 2, respectively. The coefficient of variation of the quantification value at 5 μg/L, which is less than 1/10 of the target value of 80 μg/L, was within 5%, a good result.

Figure 1. Calibration curve of Di(2-ethylhexyl)phthalate

Figure 2. SIM chromatogram of Di(2-ethylhexyl)phthalate at 5 μg/L

Table 2. Coefficient variation (C.V.) of Di(2-ethylhexyl)phthalate at 5 μg/L

Quantitation value (μg/L) C.V. %
#1 #2 #3 #4 #5
5.24 5.04 5.33 5.24 5.10 2.2
3. Dichloroacetonitrile and Chloral hydrate 3.1. Measurement

The test solution was prepared by dilution with MTBE to achieve concentrations of 1, 3, 5, and 15 μg/L of dichloroacetonitrile and chloral hydrate in the test water before pretreatment. 1, 2, 3-trichloropropane was added as an internal standard at a concentration of 125 μg/L.

3.2. Result

The calibration curves for dichloroacetonitrile and chloral hydrate are shown in Figure 3. The correlation coefficients for the relative calibration curves were more than 0.999. The SIM chromatograms and the coefficient of variation values of the quantification values for the continuous measurement of 1 μg/L samples at n=5 are shown in Figure 4 and Table 3, respectively. The coefficient of variation of the quantification value at 1 μg/L, which is less than 1/10 of the target value (dichloroacetonitrile: 10 μg/L, chloral hydrate: 20 μg/L), was within 5%, a good result.

Figure 3. Calibration curve of dichloroacetonitrile & chloral hydrate

Figure 4. SIM chromatograms of dichloroacetonitrile & chloral hydrate at 1 μg/L

Table 3. Coefficient variation of dichloroacetonitrile & chloral hydrate at 1μg/L

Compound name Quantitation value (μg/L) C.V. %
#1 #2 #3 #4 #5
Dichloroacetonitrile 1.01 1.00 1.03 1.01 1.01 0.7
Chloralhydrate 1.04 1.05 1.08 1.01 1.04 2.4
4. Summary

Di(2-ethylhexyl) phthalate, dichloroacetonitrile, and chloral hydrate were measured by the same column using a hydrogen carrier gas. The linearity of the calibration curve and the reproducibility at concentrations less than 1/10 of the target value were good, confirming that the measurement by the same column using a hydrogen carrier gas can be used adequately.

PDF (431 KB)
Solutions by field

Chemistry

Environment

Related products

JMS-Q1600GC UltraQuad™ SQ-Zeta Gas Chromatograph Quadrupole Mass Spectrometer

Attachments

  • Original Link
  • Permalink

Disclaimer

Jeol Ltd. published this content on 26 March 2024 and is solely responsible for the information contained therein. Distributed by Public, unedited and unaltered, on 26 March 2024 06:23:06 UTC.