JEOL : Peptide Analysis using Field Desorption (FD) of JMS-T2000GC AccuTOF™ GC-Alpha
March 26, 2024 at 02:24 am EDT
Share
Introduction
Field Desorption (FD) is an ionization method that utilizes the tunneling effect in a high electric field generated between an emitter coated with a sample and an electrode to which a high voltage is applied. It is known as a soft ionization method that provides only molecular weight information because the internal energy given to the sample during ionization is low and fragmentation is unlikely to occur. In FD measurements, the sample-coated emitter is introduced directly to the ion source using a dedicated probe. Therefore, it is possible to measure difficult-to-volatile compounds, polar compounds, thermally unstable compounds, etc., which are difficult to measure with GC-MS. The combination of FD and time-of-flight mass spectrometer (TOFMS) enables calculation of the molecular formula of the measured compound. An overview of FD and sampling technique in the gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha are described in MSTips No. 355 and MSTips No. 403.
In this MSTips, we report the results of the analysis of the peptide angiotensin I using FD of the JMS-T2000GC as an example of analysis of a highly polar compound.
Experimental
Commercially available angiotensin I human acetate salt hydrate was used as the sample (Figure 1). Angiotensin I was prepared to 10 mg/mL using a 0.1% TCA:methanol mixture (1:1). Then, 1 μL each of this sample and 1 mg/mL copper(II) phthalocyanine (α-type) methanol solution (used as a mass drift corrector) were applied to the emitter for measurement. A gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha was used for the measurements. An EI/FI/FD combination ion source was used, and FD was used as ionization method. Other detailed measurement conditions are shown in Table 1.
Figure 2 shows the TICC for this measurement. Two peaks were detected; the left peak was derived from angiotensin I, and the right peak was derived from copper phthalocyanine.
Figure 2 Total ion current chromatogram
Figure 3 shows the mass spectrum created from the peak of angiotensin I. The [M+H]+, [M+2H]2+, [M+3H]3+ of angiotensin I were detected (Figure 3). Table 2 shows the elemental composition estimated result of these peaks. The mass error was calculated to be within -1.5 mDa, which is a good value. Furthermore, the isotope pattern of [M+H]+ was roughly the same as the simulation result (Figure 4), indicating that good measurement results were obtained. Essentially, peptides cannot be measured using a GC-MS due to its high polarity. But FD which is a direct MS method can be used for peptides analysis, then molecular formula of the peptide can be easily calculated using a FD-TOFMS system.
Figure 3 Mass spectrum of angiotensin I
Ion Type
Measured m/z
Formula
Calculated m/z
Mass Error [mDa]
[M+H]+
1296.68462
C62 H90 N17 O14
1296.68477
-0.15
[M+2H]2+
648.84471
C62 H91 N17 O14
648.84602
-1.31
[M+3H]3+
432.89845
C62 H92 N17 O14
432.89977
-1.32
Figure 4 Isotope pattern simulation result of [M+H]+
Conclusions
In this MSTips, we introduced the analysis results of angiotensin I as an analysis example of highly polar compound using FD. As a result of the analysis, it was confirmed that protonated molecule related to molecular weight could be observed, and furthermore, accurate mass could be obtained with good accuracy. These results indicate that FD can be used to analyze highly polar compounds that are difficult to measure by GC-MS.
PDF (472 KB)
Solutions by field
Chemistry
Food / Plant
Biology / Life science
Related products
JMS-T2000GC AccuTOF™ GC-Alpha High Performance Gas Chromatograph - Time-of-Flight Mass Spectrometer
Attachments
Original Link
Permalink
Disclaimer
Jeol Ltd. published this content on
26 March 2024 and is solely responsible for the information contained therein. Distributed by
Public, unedited and unaltered, on
26 March 2024 06:23:06 UTC.
JEOL Ltd. is a Japan-based company mainly engaged in the manufacture, sale, maintenance, development and research of scientific instruments, semiconductor related equipment, industrial and medical equipment, and related products and parts. It operates through three business segments. Science and Measurement Equipment segment manufactures and sells three types of equipment. Electron optical devices include electron microscopes, peripheral devices and others. Analytical instruments include nuclear magnetic resonance equipment, electron spin resonance equipment and others. Measurement and inspection equipment include scanning electron microscopes, analytical scanning electron microscopes, peripheral equipment and others. Industrial Equipment segment manufactures and sells electron beam lithography equipment, linear electron guns and power supplies and others. Medical Equipment segment manufactures and sells clinical test information processing systems, and automatic amino acid analyzers.